Specifications

Recovery time

Sampler temperature recovers to 4 °C within 5 minutes after the door has been
held open for one minute in 24 °C (75 °F) ambient environment while in an active
cooling cycle.

Pull down time

Air temperature drops from 24 °C (75 °F) to 4 °C (39 °F) within 20 minutes
(typical).

Certifications

North America: cETLus listed -Conforms to UL 61010-1, Certified to CSA C22.2
No. 61010-1 and UL 471 and CSA C22.2 No. 120.
Europe / IEC: CE - EN / IEC 61010-1 and EN / IEC 60335-2-89 (safety), EN / IEC
61326 (EMC) & CISPR 11 (RF emissions)

Sample bottle capacity (see

Figure 3 on page 11

and

Figure 4 on page 12

)

Single bottle mode

(1) 21 L (5.5 gal) polyethylene or (1) 10 L (2.5 gal) polyethylene or (1) 10 L (2.5
gal) glass bottle

Multiple bottle mode

(2) 10 L (2.5 gal) polyethylene and/or 10 L (2.5 gal) glass bottles
(4) 10 L (2.5 gal) polyethylene and/or 10 L (2.5 gal) glass bottles
(8) 2.3 L (0.6 gal) polyethylene and/or 1.9 L (0.5 gal) glass bottles
(24) 1 L polyethylene and/or 350 mL glass bottles

Intake accessories

Strainers

316 stainless steel in standard size, high velocity or low profile for shallow depth
applications and Teflon

®

/316 stainless steel in standard size

Sample intake tubing

9.5 mm (

3

/

8

in.) I.D. vinyl or Teflon

®

-lined polyethylene

SD900 controller

Enclosure

High-Impact, injection-molded PC/ABS blend; submersible, watertight, dust-tight,
corrosion, and ice resistant; NEMA 4X, 6, IP 67

Power requirements

Refrigerated sampler: 15 VDC provided by 8754500 power supply
AWRS sampler: 15 VDC provided by integral power supply

Overload protection

7 amp DC line fuse for pump

Pump

Peristaltic high speed, with spring-mounted rollers

Pump rollers

Nylatron,

impact/corrosion resistant

Pump enclosure

Track is injection-molded polyphenylene sulfide. Cover is polycarbonate, high
impact resistant. Pump enclosure rated IP37.

Pump tubing

9.5 mm I.D. x 15.9 O.D. mm (

3

/

8

in. x

5

/

8

in.) silicone

Pump tubing life

20,000 sample cycles under the following conditions:

1 L sample volume
1 rinse
6 minute pacing interval
16 ft of

3

/

8

in. intake tube

15 ft of vertical lift
70 °F sample temperature

Tubing replacement time

< 1 minute using pre-cut pump tube

Maximum vertical lift to draw sample

Minimum of 28 ft, using 29 ft of

3

/

8

-in. vinyl intake tube at sea level at 20–25 °C

(68–77 °F)

Pump flow rate

1.25 gpm (4.8 L/min) at 3 ft (1 m) vertical lift using

3

/

8

-in. intake tube

Typical sample volume repeatability

± 5% of 200 mL sample volume using uncalibrated liquid detect with 15 feet
vertical lift,16 feet of

3

/

8

-in. vinyl intake tube configured for a single bottle using full

bottle shut-off at room temperature and 5000 ft elevation

Typical sample volume accuracy

± 10% of 200 mL sample volume using uncalibrated liquid detect with 15 feet
vertical lift,16 feet of

3

/

8

-in. vinyl intake tube configured for a single bottle using full

bottle shut-off at room temperature and 5000 ft elevation