3 status for stormwater program, 1 status of sdi-12 measurements, 4 sample history – Hach-Lange SIGMA SD900 Refrigerated Sampler and All Weather Refrigerated Sampler User Manual User Manual

Page 55

Advertising
background image

53

Sampler operation

4.7.3 Status for stormwater program

The status screen shows information about the stormwater
program similar to that shown for the main program (Ready,
Running, Halted or Complete).

Use the

RIGHT

ARROW

key to access additional information about

the program.

4.7.3.1 Status Of SDI-12 Measurements

If a sonde is detected on the SDI-12 interface and the sonde
provides measurement data, the user can view the current status of
these measurements using the 'STATUS' hard key or by selecting
STATUS in the main menu.

The measurement status screen displays current measurement of
each channel on the sonde including the measurement name and
unit assigned to each specific channel. The measurements will be
updated based on the user-defined scan interval (see

SDI-12

configuration on page 60

). Navigation arrows on the screen will

indicate whether additional channel information is available.
CH0..CHX indicates the current position within the available
channels.

4.7.4 Sample history

Sample history displays the sample number, bottle number, result
and time/date for each sample. The following results can be shown:

Success—the sample was drawn successfully

Bottle full—the full bottle shut-off was activated

Rinse error—an error occurred during the rinse cycle

User abort—a user pressed the

STOP

key to end the

sample cycle

Arm faulty—the distributor arm did not move correctly

Pump fault—a fault occurred when the pump was operating

Purge fail—an error occurred during the purge cycle

Sample time out—liquid not detected within the timeout period

Power fail—a power failure occurred during sampling

If a program is running or has been halted, the sample history is
displayed for the currently running sample program. If the program
has been completed, the sample history is displayed for the most
recently completed sampling program. The sample history is
automatically erased when a new program is started.

Procedure:

1. Select

DIAGNOSTICS

>

SAMPLE HISTORY

from the main menu

to view the sample history.