7.12.3 SCREEN OPERATION

3-1.

TO adjust the calibration, a simple calculation must be performed. With the master
torque transducer in-line with the FEC torque transducer, perform 10 fastenings and
collect the fastening torque data from the FEC and the master transducer.

3-2.

DETErmine the average reading from both transducers and insert them into the
following formula:


(AVG. OF MASTER) - (AVG. OF FEC)
---------------------------------------------------- X 100 = RATIO [%] preset
AVERAGE OF MASTER


EXAMPLE:

FEC AVERAGE = 10.0 KGM MASTER AVERAGE = 9.8 KGM

9.8 KGM - 10.0 KGM
------------------------------ X 100 = -2.0408 (RATIO [%])
9.8 KGM

3-3.

The calibration adjustment range is +/- 9.9%. The normal value to set the calibration
percentage is 0%. THIS value should change only if transducer matching is not
possible. The calibration percentage should be as close as possible to 0%.

3-4.

TO change the calibration ratio %, set the RUN/PROGRAM key switch to PROGRAM
and depress the [SEL] key. THE screen will highlight the first spindle RATIO [%]

3-5. USINg

the

[↑]

[↓]

keys, move the highlighted cursor to the desired spindle.

3-6.

USING the numerical,

[±]

and [.] keys, enter the RATIO [%] value and depress the

[↵]

key. THe screen will change to display the new data.

3-7.

AFTer all data has been entered, depress the [SEL] key and the question "DATA
CHANGE OK?" will appear. SELect YES if the data is to change or NO if the old
data should remain.

3-8.

TO print out the torque calibration data, use the

[←]

[→]

keys to move the cursor on

the menu bar to [PRINT] and depress the

[↵]

key.

NOTE: After adjusting the Calibration Value, the machine configuration should be saved to disk.







Chapter 7: System Operation

Page 7-76