A

PPENDIX

A

S

PECIFICATIONS

A–16 www.zoll.com

9650-0350-01 Rev. J

Arrhythmia performance is reported according to the article, Kerber RE, Becker LB,
Bourland JD, Cummins RO, Hallstrom AP, Michos MB, Nichol G, Ornato JP, Thies WH,
White RD, Zuckerman BD. “Automated External Defibrillators for Public Access
Defibrillation: Recommendations for Specifying and Reporting Arrhythmia Analysis
Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety,” Circ J Am
Heart Assoc.
1997;95:1677-1682.

References

Young KD, Lewis RJ. “What is confidence? Part 2: Detailed definition and determination of
confidence intervals”. Ann Emerg Med. September 1997;30:311-318.

Beyer WH. Percentage Points, F-Distribution Table. CRC Standard Mathematical Tables.
28

th

ed. Boca Raton, Fla: CRC Press; 1981:573.

Table A-5.

Clinical Performance Results (Pediatric Patients)

Rhythms

Sample

Size

(9 second

records)

Performance

Goals

Observed

Performance

90% One-sided

Lower

Confidence

Limit

Shockable
(49 patients)

Sensitivity

Coarse VF

42

>90%

100% (42/42)

93.1%

Rapid VT

82

>75%

93.9% (77/82)

87.6%

Nonshockable
(155 patients)

Specificity

NSR

208

>99%

100% (208/208)

98.6%

AF, SB, SVT

a

, heart block,

idioventricular, PVCs

a. 161 of the 348 abnormal rhythm records were SVT (72 patients). The SVT heart rates ranged

from 152 to 302 beats per minute.

348

>95%

99.4% (346/348)

98.2%

Asystole

29

>95%

100% (29/29)

90.2%

Intermediate
(16 patients)

Fine VF

0

Report only

Other VT

40

Report only

90% (36/40)

78.6%