Data storage, Device history, Patient clinical data – ZOLL AED Pro Rev J User Manual

Page 75

Advertising
background image

Data Storage

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

6–3

Data Storage

The AED Pro unit stores device history and patient clinical data in nonvolatile memory.

The unit retains the device history and clinical data even when powered off or when the battery
pack is removed. Clinical data is erased only when the device is powered on and electrodes are
attached to a new patient. If configured to do so, the unit can store data for more than one
patient.

Device History

The AED Pro unit keeps a log of its status information, including:

Unit model name

Device serial number

Hardware revision number

Application software and boot code revision numbers

Language file version number

Total shocks delivered

Battery life status (percentage of charge remaining)

Ready indicator status

Elapsed time since installation of the battery

Date and results of last self-test

Error log

To view device history, use ZOLL Administration Software.

Patient Clinical Data

During clinical mode, the device stores the following, with date and time notation:

Electrode attachment

Electrode type (CPR-D•padz, CPR stat•padz, stat•padz II, or pedi•padz II defibrillation
pads, or AED Pro ECG cable)

Results of ECG analysis

Continuous ECG data

Number of shocks delivered

Energy delivered

Patient impedance

Voice prompts

CPR data (depth and rate of chest compressions)

To view and analyze patient clinical data, use ZOLL RescueNet Code Review software.