C–4 www.zoll.com

9650-0350-01 Rev. J

Breathing Prompt Delay

Sets the period over which the unit issues the check
breathing prompts.

•

2 seconds

•

4 seconds

•

6 seconds — default

•

9 seconds

•

15 seconds

•

20 seconds

Responsiveness/Patient

Selects the wording for this prompt.

•

OFF — default

•

Check Responsiveness

•

Check Patient

Responsiveness/Patient Prompt Delay

Sets the duration before the next prompt or disables
this prompt.

•

2 seconds

•

4 seconds — default

•

6 seconds

•

8 seconds

Circulation/Pulse

Selects the wording for this and related prompts.

•

OFF — default

•

Check Circulation

•

Check Pulse

Circulation/Pulse Prompt Delay

Sets the duration before the next prompt or disables
this prompt.

•

2 seconds

•

5 seconds

•

7 seconds

•

10 seconds — default

•

15 seconds

•

20 seconds

Press Treatment/Shock

Selects the wording for this and related prompts.

•

Press Flashing Treatment Button

•

Press Flashing Shock Button —
default

CPR

CPR Monitoring

When this option is checked (On) and CPR-D•padz or
CPR stat•padz are attached, the unit performs CPR
monitoring and prompts the rescuer.

•

On — default

•

Off

No-Shock CPR Period

Sets the duration of the CPR period following a
no-shock result for the first analysis in a stack.

When you select Extended, the device remains in the
CPR state indefinitely and displays the ANALYZE
softkey. You must press the ANALYZE softkey to move
to analysis.

•

30 seconds

•

60 seconds

•

90 seconds

•

120 seconds — default

•

150 seconds

•

180 seconds

•

Extended

Cycle through CPR

When this option is checked (On) and pads are
detached from the patient for more than one and
one-half minutes, the unit prompts the rescuer to
perform CPR.

•

On — default

•

Off

Configurable Option

Possible Values