Guidance and Manufacturer’s Declaration —
Electromagnetic Emissions

The ZOLL AED Pro device is intended for use in the electromagnetic environment specified
below. The customer or operator should ensure that the device is used in such an environment.

Table A-1.

EMC Specifications

Emissions Test

Compliance

Electromagnetic Environment –

Guidance

RF emissions

CISPR 11

Group 1

The ZOLL AED Pro unit uses RF energy
for its internal function only. Therefore, its
RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.

RF emissions

CISPR 11

Class B

Harmonic emission

IEC 61000-3-2

Not applicable

Voltage fluctuations/
flicker emission

IEC 61000-3-3

Not applicable

Medical electrical equipment needs special precautions regarding EMC and needs to be
installed and put into service according to EMC information provided in this document.