Messages in Manual Mode

During manual mode, the following messages can be displayed:

Message

Description

PLUG IN CABLE

The unit started up without an electrode cable plugged
in. Plug the cable into the unit.

CHECK DEFIB PADS

The defibrillation electrode pads became disconnected.

Check the pads to ensure that they are properly applied
to the patient, and verify that the cable is undamaged
and plugged into the unit.

CHECK PATIENT

Evaluate the status of the patient. The unit detected
either a shockable rhythm or a low heart rate.

DON’T TOUCH PATIENT, CHARGING

The defibrillator is charging.

Do not touch the patient.

RELEASE SHOCK BUTTON

The Shock button was pressed while the unit was
charging.

Do not press the Shock button until the unit emits the
charge-ready tone, and the button begins flashing.

RELEASE LEFT SOFTKEY

or

RELEASE RIGHT SOFTKEY

A softkey was pressed for more than 10 seconds; the
unit automatically switched to semiautomatic mode.

Release the softkey.