Optional Maintenance for Technical Professionals

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

7–7

Optional Maintenance for Technical Professionals

The AED Pro is calibrated at the factory and requires no testing other than the self-tests it
performs. Qualified professionals who want to perform additional tests can use the following
procedure.

Required equipment

•

AED Pro simulator (or equivalent)

Before you begin

Be sure the unit and the simulator/tester are turned off.

Procedure

To test the unit:

Step

Action

1

Connect the AED Pro simulator to the unit’s patient cable connector.

2

Power on the simulator and the AED Pro unit.

3

Verify that all of the following occur:

•

The Ready indicator initially displays a red “X”, which changes to a green check within
10 seconds after the unit is turned on.

•

The unit issues the voice and text message UNIT OK within 10 seconds (if configured).

•

The screen displays the shock count and the elapsed time.

4

Set the simulator to send a VF rhythm to the AED Pro unit.

5

After the sequence of patient assessment prompts, verify that the unit does the following:

•

Issues the voice and text prompt DON’T TOUCH PATIENT, ANALYZING.

•

Analyzes the ECG rhythm.

•

Issues the voice and text message SHOCK ADVISED.

•

Charges the defibrillator.

•

Issues the voice and text prompts DON’T TOUCH PATIENT and PRESS FLASHING
SHOCK BUTTON
.

6

Verify that the AED Pro unit sounds the charge-ready tone and that the Shock button
flashes repeatedly.

7

Press the Shock button. Verify that the simulator indicates that a shock was delivered and
that the unit updates the displayed shock count.

Note: The simulator can verify the unit’s ability to deliver energy but cannot verify that the

correct energy was delivered. To verify the level of delivered energy, use a
defibrillator analyzer and universal adapter cable in place of the simulator.

8

Immediately after shock delivery, change the simulator to send a normal sinus rhythm
(NSR) to the AED Pro unit.

9

Verify that the AED Pro unit performs a new rhythm analysis, resulting in the message NO
SHOCK ADVISED
, followed by the voice and text prompt:

START CPR