Switching to Manual Mode

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

4–3

Switching to Manual Mode

When the unit is operating in semiautomatic mode, you can override the automated functions
and switch to manual operation. Changing modes does not affect the preselected energy level.

Before you begin

Be sure the defibrillation electrode pads are properly attached to the patient and are plugged
into the unit. (For more information, refer to Chapter 3.)

Procedure

To override semiautomatic mode and begin manual mode:

To return to semiautomatic mode after manual override, turn the unit off and then back on
again.

Step

Action

Result

1

Press and hold both (unlabeled) softkeys
simultaneously for at least 3 seconds.

The MANUAL and SEMI-AUTO softkey
labels appear.

2

To change to manual mode, press the
MANUAL softkey.

Or to remain in semiautomatic mode, press
the SEMI-AUTO softkey.

Note: If you do not press either softkey

within 10 seconds, the unit reverts
to semiautomatic mode.

The unit changes to manual mode, and
displays the CHARGE softkey label.