Heart rate averaging

The AED Pro averages the interval between the last 5 detected beats.
On startup, the AED Pro averages the rate between detected beats
once two beats are detected, until a full 5 beats have been received.
The rate is updated every beat. After this condition is met, the meter is
updated every beat with an average of the last 5 beats.

If a period of time greater than 5 seconds elapses without a beat
detected, the heart rate meter reports a rate of 0 bpm, which is
repeated every 5 seconds.

Accuracy and response
time to irregular rhythm
per EN 60601-2-27
2nd Ed.: 2005

Ventricular Bigeminy (Figure A1) - 40 bpm

Slow alternating ventricular bigeminy (Figure A2) - 55-65 bpm

Rapid alternating ventricular bigeminy (Figure A3) - 59-60 bpm

Bidirectional systoles (Figure A4) - 59-60 bpm

Response time to
change in heart rate

80 to 120 bpm: 4 seconds
80 to 40 bpm: 4 seconds

Time to alarm for
tachycardia per
EN 60601-2-27
2nd Ed.: 2005

(Figure B1) 206 bpm (1 mV): 7.9 seconds

206 bpm (.5 mV): 8.7 seconds
206 bpm (2 mV): 8.2 seconds

(Figure B2) 195 bpm (2 mV): 7.9 seconds

195 bpm (1 mV): 7.1 seconds
195 bpm (4 mV): 8.0 seconds

ECG leads off sensing

A DC current of < 10

A per lead wire is supplied to the patient.

Defibrillation electrodes
leads off sensing

67 kHz square wave, <1 mA

Data Recording and Storage

Type

Nonvolatile memory

Capacity

5.8 hours of ECG data and CPR data

20 minutes of audio recording and ECG data if Audio Recording
Option is enabled

Display Screen

Display type

Liquid crystal display (LCD)
High resolution, 320 pixels by 240 pixels

Viewable area
(height width)

2.27 in 3.02 in
5.76 cm 7.68 cm

Sweep speed

25 mm/s ±5%

Viewing time

2.96 seconds (if CPR gauge displayed)
3.2 seconds (without CPR gauge)