Semiautomatic Defibrillation

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

3–11

WARNING! You have 30 seconds to perform the following step, otherwise the

defibrillator automatically disarms itself. (During the final 10 seconds, the
charge-ready tone sounds intermittently to indicate that time is expiring.)

Before discharging the defibrillator, warn everyone to STAND CLEAR.
Verify that no one is touching the patient, the bed rails, or any other
potential pathway for electrical current.

6

Press and hold the Shock button until
treatment is delivered.

The unit delivers the shock and updates
the shock count.

Depending on the number of shocks that
have been delivered, the unit either
resumes ECG analysis or prompts you to
perform CPR.

7

Does the unit resume ECG analysis?

— If NO, continue with step 8.

— If YES, return to step 3.

When the unit resumes ECG analysis, it
displays the message:

DON’T TOUCH PATIENT, ANALYZING

8

Follow the prompts to perform CPR, until
directed to stop.

(CPR-D•padz or CPR stat•padz only) After
the first few chest compressions, the
metronome begins beeping. Try to time each
compression with the metronome beep.
Check the on-screen gauge to ensure that
the compression depth is adequate.

The unit displays the message:

START CPR

At the end of the defined CPR period, the
unit displays the message:

STOP CPR

Note: If CPR-D•padz or CPR stat•padz are
attached, the unit also monitors the rate
and depth of chest compressions and can
issue these related voice and text prompts:

PUSH HARDER

GOOD COMPRESSIONS

In addition, if the unit does not detect chest
compressions, it issues the following voice
and text prompt every 10 seconds (default):

CONTINUE CPR

9

When the unit resumes ECG analysis, return
to step 3.

During ECG analysis, keep the patient
motionless, and do not touch the patient.

Step

Action

Result

ZM040011A