Applying ECG Electrodes

Proper application and placement of electrodes is essential for ECG monitoring. Good contact
between the electrode and the skin minimizes motion artifact and signal interference. ZOLL
recommends the use of high-quality Ag/AgCl (silver/silver chloride) ECG electrodes.

Before you begin

Check the electrodes to ensure they are free from damage and are not past the expiration date
on the packaging.

Procedure

To apply ECG electrodes to the patient:

Step

Action

1

Identify the appropriate sites for electrode placement
(refer to “ECG Electrode Placement” on page 5-3).

2

Clean and abrade the patient’s skin to remove the outer
layer of dead tissue.

Ensure that the skin at the electrode site is dry.

3

Snap the leads onto each electrode.

Check for good contact between the electrode and the
lead termination.

4

Peel the protective backing from the ECG electrode.

Keep the adhesive surface free of electrolyte gel.

WARNING! Do not use electrodes if the gel is dried or

damaged.