Descriptions of AED Pro Configurable Settings

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

C–3

Shock Energy Level and Charging

Adult First Shock Energy

Sets the energy level in joules for the first shock for an
adult patient.

120 J — default

150 J

200 J

Adult Second Shock Energy

Note: This value cannot be less than the value

selected for the first adult shock.

120 J

150 J — default

200 J

Adult Third Shock Energy

Note: This value cannot be less than the value

selected for the second adult shock.

120 J

150 J

200 J — default

Pediatric First Shock Energy

Sets the energy level in joules for the first shock for a
pediatric patient.

50 J — default

70 J

85 J

Pediatric Second Shock Energy

Note: This value cannot be less than the value

selected for the first pediatric shock.

50 J

70 J — default

85 J

Pediatric Third Shock Energy

Note: This value cannot be less than the value

selected for the second pediatric shock.

50 J

70 J

85 J — default

Enable Lay Rescuer

When this option is checked (On), the unit issues the
following voice and text prompts after completion of the
power-on self-test and entry into clinical mode:

STAY CALM

CALL FOR HELP

Off — default

On

Jump to Analysis

When this option is checked (On), the unit immediately
begins ECG analysis when pads are attached to the
patient, except during CPR periods.

Note: The Start with CPR option, if also checked (On),

overrides this setting.

On — default

Off

Unit OK

When this option is checked (On), the unit issues the
message Unit OK after a successful power-on self-test.

On

Off — default

Clinical Protocol Voice Prompts

Breathing

Allows you to select a check breathing prompt, based
on the guidelines you want to follow.

When disabled (Off), these prompts are not issued.

OFF — default

Open Airway, Check Breathing,
Give Two Breaths

Open Airway, Check Breathing

Give Two Breaths

Configurable Option

Possible Values