COMMAND DETAILS

71

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:FALL?
Selects Channel 1 as the source, and then measures
the fall time.

:MEASure:PWIDth

Description

Returns the first positive pulse width.

Syntax

:MEASure:PWIDth?

Related
Commands

:MEASure:SOURce<X>

Return parameter <NR3>

Returns the positive pulse width.

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:PWIDth?
5.0E-6
Selects Channel 1, and then measures the positive
pulse width.

:MEASure:NWIDth

Description

Returns the first negative pulse width timing.

Syntax

:MEASure:NWIDth?

Related
Commands

:MEASure:SOURce<X>

Return parameter <NR3>

Returns the negative pulse width in
seconds.