Measurement:meas<x>:type, Measurement:meas<x>:value – GW Instek GDS-200 Series Programming Manual User Manual

Page 82

Advertising
background image

GDS-200 & GDS-300 Programming Manual

82

Example

:MEASUrement:MEAS1:SOURCE1
CH1

Returns the (first) source for measurement 1.

:MEASUrement:MEAS<X>:TYPe

Description

Sets or queries the measurement type for a selected
automatic measurement. This is a statistics related
command.

Syntax

:MEASUrement:MEAS<X>:TYPe
{PK2pk | MAXimum | MINImum | AMPlitude | HIGH |
LOW | MEAN | CMEan | RMS | CRMs | AREa | CARea |
ROVShoot | FOVShoot | RPReshoot | FPReshoot |
FREQuency | PERIod | RISe | FALL | PWIdth | NWIdth |
PDUTy | PPULSE | NPULSE | PEDGE | NEDGE
| FRRDelay | FRFDelay | FFRDelay | FFFDelay |
LRRDelay | LRFDelay | LFRDelay | LFFDelay | PHAse
| ?}

Related
commands

:MEASUrement:MEAS<X>:SOURCE<X>

Parameter

MEAS<X>

The automatic measurement number
from 1 to 4.

Return parameter Returns the measurement type

Example

:MEASUrement:MEAS1:TYPe RMS
Sets measurement 1 to RMS measurement.

:MEASUrement:MEAS<X>:VALue

Description

Returns the measurement results for the selected
measurement. This is a statistics related command.

Syntax

:MEASUrement:MEAS<X>:VALue?

Related
Commands

:MEASure:SOURce<X>
:MEASUrement:MEAS<x>:TYPe

Set

Query

Query

Advertising
This manual is related to the following products: