72

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:NWIDth?
4.995E-04
Selects Channel 1, and then measures the negative
pulse width.

:MEASure:PDUTy

Description

Returns the positive duty cycle ratio as percentage.

Syntax

:MEASure:PDUTy?

Related
commands

:MEASure:SOURce<X>

Return parameter <NR3>

Returns the positive duty ratio.

Chan Off

Indicates the source channel is not
activated.

Note

Before using this command, select the
measurement channel. See the example below.

Example

:MEASure:SOURce1 CH1
:MEASure:PDUTy?
5.000E+01
Selects Channel 1, and then measures the positive
duty cycle.

:MEASure:PPULSE

Description

Returns the number of positive pulses.

Syntax

:MEASure:PPULSE?

Query

Query