Jenway Genova Nano German User Manual

Page 104

Advertising
background image

Err 4

Microswitch Failure (service only)

This error indicates that the microswitch was not found.

The most likely causes of this error are:

1. The microswitch is broken.

Solution:

Contact a service technician.

Err 5

Light Saturation Not Found

This error indicates that the peak light hasn’t been found at
zero.
The most likely causes of this error are:
1. Lamp failure.
2. Deteriorating lamp signal.
3. Sample or cuvette in the sample holder.
Solution:
Ensure the sample holder is empty. Restart the unit, if the
problem persists contact a service technician.

Err 7

USB Module Failure

This error indicates the instrument is unable to detect a
response from the UBS module.
The most likely causes of this error are:
1. USB port failure.
Solution:
Press the key adjacent to the tick icon to carry on using the
instrument.

Restart the unit, if the problem persists contact a

service technician.

Err 8

Unable to find Vane on Turret

This error indicates that the vane position zero on the turret
cannot be found.
The most likely causes for this error are:
1. The turret carousel has been removed and not replaced.
Solution:
Check if the turret is in the sample chamber and inserted
properly. Press the key adjacent to the back icon to try
rechecking.