Jenway Genova Nano German User Manual

Page 50

Advertising
background image

9.2.1.8 Selecting a Wavelength

To adjust the wavelength press the key adjacent to the
wavelength icon. This will open the number entry screen.
Use the keys at the bottom of the screen to select the digit
to be changed, pressing the key twice to select the second
digit. The keys adjacent to the arrow icons can be used
to increase or decrease the number. Once the required
wavelength has been entered press the key adjacent to
the tick icon to save the changes and return to the settings
menu screen.

9.2.1.9 Setting the Kinetics Measurement Time

This function enables the measurement time for the
kinetics scan to be set. To access this function press the
key below the time displayed on the far right of the x-axis;
this will open a number entry screen. Use the keys at the
bottom of the screen to select the digit to be changed
and use the keys adjacent to the arrow icons to increase
or decrease the number. Once the required measurement
time in seconds has been entered press the key adjacent to
the tick icon to save and return to the settings menu.

9.3

CALIBRATION

The calibration must be performed at the same wavelength
at which the sample will be measured. Insert a cuvette
containing the blank solution into the sample chamber and
close the instrument lid. Press the key below the calibrate
to zero absorbance icon. The instrument will calibrate to
0.000 absorbance.

Once the calibration is complete the start kinetics measurement icon will be displayed and a sample can
be measured. If the wavelength is adjusted before the sample is measured the start kinetics measurements
icon will disappear and the instrument must be calibrated again at the new wavelength.

9.4

SAMPLE MEASUREMENT

Insert a cuvette containing the sample to be analysed
into the sample chamber and press the key below the
start kinetics measurements icon. If a start on level or
lag time has not been set the instrument begins taking
a photometric reading every second. As each reading is
taken it is plotted on the graph.

If a lag time has been set a progress bar will be displayed
showing the lag time counting down. Once the lag time is
complete the instrument will begin taking readings.

48

Advertising