When the accessory is fitted the instrument will look like
this.

18.2.3.3 Sipper pump

For this accessory as well as removing the passive accessory
base plate, the front panel of the instrument must also be
removed. Loosen screws 5 and 6 until the front panel can
be lifted out in the forwards direction.

Take the sipper base plate. Connect the power supply in
the bottom of the sample chamber to the connector on
the underside of the base plate. Place the base plate in the
sample chamber. Replace screws 1 to 4. Take the sipper
front panel and slot into place before retightening screws
5 and 6.

The tubing should be connected depending on the function
that the sipper pump is going to perform. All tubing must
be kept as short as possible and the tubing must not be
allowed to obstruct the ligth path.

5

6

Bi directional

flow A

(sipping)

Bi directional

flow B

(pumping)

Continuous

flow