Curve fit algorithms

Quadratic, quadratic through zero, linear, linear

through zero, interpolate

Multi-wavelength

Range

0 to 9999

Data points

Up to 4 wavelengths

Calculations

Sum, product, ratio, difference

Kinetics

Measurement Time

2 to 9999 seconds

Calibration

Blank with a single standard or factor

Display

Graphical and concentration

Analysis

Concentration, rate of change, initial and final

absorbance/%T

Resolution

Selectable 1/0.1/0.01/0.001

Spectrum/Purity Scan

Range

198 to 1000nm

Scan interval

Selectable 1, 2 or 5nm

Analysis

Absorbance or % transmittance and peak and valley

wavelengths

Multi-wavelength Plus

Range

0 to 9999

Data points

3 wavelengths + optional reference wavelength

Calculations

concentration, ratio

DNA

Measurement modes

dsDNA, ssDNA, RNA, Oligonucleotides,

260/280, 260/230, Variable Ratio

Protein

Measurement modes

Pierce 660, BCA, Bradford, Lowry, Biuret, Direct UV

OD 600

Range

0.00 E-19 to 9.99 E+19

Calibration

Blank with a single standard or factor

Units

cells/ml

Factor

0.01 E-19 to 9.99E+19

Standard

0.01 E-19 to 9.99E+19

Instrument Factor

0.001 to 9999.999

Other

Beam height

15mm

Light source

Xenon lamp

GLP

Current time and date, user ID, settings lock and

method lock

Number of users

999

Methods memory

312 (including pre-programmed methods)

Results memory

Limited by attached mass storage device

Removable media

USB (supplied)

Outputs

USB, Analogue, RS232, Internal printer

Power

24V

Size (w x d x h)

275 x 400 x 220mm

Weight

6kg

*Assessment must be performed with a 630 204 - 10 x 10mm path length cuvette holder installed.