specifically. This function can be accessed from the statistics menu or from the operating menu by
pressing the key adjacent to the kinetics moving line icon.

In the kinetics moving line screen the initial and final time
lines appear as two vertical lines. The line which is selected
is represented by a dashed line ---- and can be moved
using the keys below the greater than (>) or less than (<)
icons at the bottom of the screen.

Move the dashed line to the point on the x-axis which
is to be the initial time. To select the final time press the
key adjacent to the toggle icon. This will change the initial
line from a dashed line to a solid line. The final time line
will change from a solid line to a dashed line. Use the
keys below the greater than (>) or less than (<) icons to
move the line to the final time on the x-axis by reducing or
increasing in 1 second time intervals.

The double greater than (>>) or double less than (<<) icons move the line in 5 second time intervals.

Once the required initial and final times have been selected
press the key adjacent to the S icon to display the statistics
menu for that portion of the experiment. The information
displayed includes concentration, absorbance per minute
(rate of change), correlation coefficient (r

2

), start time

absorbance and end time absorbance. If the operating
mode selected was % transmittance then the statistics
will be % transmittance instead of absorbance. To exit the
statistics menu press the key adjacent to the tick icon.