Jenway Genova Nano German User Manual

Page 97

Advertising
background image

This opens the Peltier settings screen. Use the keys at the
bottom of the screen to select the digit to be changed and
use the arrow icons to increase or decrease the number.
The temperature can be set in °C or °F by pressing the key
adjacent to the °C icon.

Repeat presses will cycle between °C and °F. Once the
required temperature has been selected press the key
adjacent to the tick icon to save and return to the operating
menu. The Peltier will begin to heat or cool depending on
the current temperature.

18.3.3 Sipper pump

When the sipper is in use the sipper pump icon is displayed
in the bottom right hand corner of the screen. The sipper
pump can operate in manual or timed mode, depending
on the option selected in sipper pump settings. If the
manual mode is selected an arrow icon indicating pump
direction will be displayed below the sipper pump icon.

If the timed mode is selected a clock icon will be displayed
adjacent to the sipper pump icon.

To open the sipper pump settings hold the key below the
sipper pump icon for 2 seconds.

18.3.3.1 Manual Sipper Pump Settings

To operate the sipper pump in manual mode press the key
adjacent to the manual sipper icon. Select the preferred
pump direction by pressing the key below the forwards or
backwards arrow. Press the key adjacent to the tick icon to
save and return to the expanded operating menu.

To perform a measurement place the sipper tubing into the
sample and press the key below the sipper pump icon.