Jenway Genova Nano German User Manual

Page 105

Advertising
background image

20.2

TOUBLESHOOTING GUIDE

Issue

Solution

Unable to achieve zero absorbance or
100% transmittance when calibrating

Unable to achieve a reading when
measuring a sample

Unable to add additional points to the
spectral points analysis table

Unable to print results using the internal
printer

Unable to save results to the USB
memory stick

Unable to continue kinetics measurement
after cancelling abort action


The peaks and valleys table icon is not
displayed following a spectrum scan

The measure sample icon disappears after
changing the wavelength

Unable to save methods

20.3

TECHNICAL SUPPORT

Jenway have a dedicated Technical Support team made up of experienced scientists who are on hand to
help with any applications advice and questions you may have about our products and how to use them.
If you require any technical or application assistance please contact the team at:

E-mail: [email protected]
Phone: +44 (0)1785 810433
Fax: +44 (0)1785 810405

Ensure that there is not a sample in the sample chamber.
Ensure the instrument lid is closed before and during the
calibration.
Ensure the lamp is working – if the lamp has failed please
request service assistance.

Ensure the correct cuvette is being used so that light isn’t
being absorbed by the cuvette.
Ensure the sample isn’t too dense that light is not
transmitted through the sample.
Ensure the lamp is working.

The table can only hold 6 points. Delete one or more
points before selecting additional points to add to the
table.

Ensure internal printer selected in autolog menu.
Ensure there is paper in the unit.
Ensure there is a result displayed on the screen.

Ensure the USB stick is inserted into the front of the unit.
Ensure the USB memory stick has free capacity.
Ensure there is a result displayed on the screen.

Due to the nature of the kinetics measurements stopping
the measurement before completion will result in some
measurement time being lost. Therefore the experiment
cannot be continued. It must be restarted.

The peaks and valleys threshold is turned off. Change
this value to 1%, 5% or 10% and the peaks and valleys
table icon will be displayed.

A calibration must be performed at the new wavelength.
When the calibration is complete the measure sample
icon will be displayed.

Ensure the method lock is not active.
Ensure the method is not a pre-loaded locked method.