This will expose the bottom of the sample chamber with
the power supply connection needed to operate the active
accessories.

18.2.3.1 Automatic 8 cell turret

Take the 8 cell turret base plate. Connect the power supply
in the bottom of the sample chamber to the connector on
the underside of the base plate. Place the base plate in the
sample chamber. Replace screws 1 to 4.

Take the 8 cell carousel and place on top of the motor,
taking care to align the three ball bearings with the grooves
on the motor shaft. Gently push the carousel down onto
the motor shaft until it is located into place. Gently rotate
the carousel until there is some resistance. The carousel is
now in the correct position.

If the fitting is too tight use a small screw driver to loosen
the ball bearings before pushing the carousel down onto
the shaft.

18.2.3.2 Peltier

For this accessory as well as removing the passive accessory
base plate, the front panel of the instrument must also be
removed. Loosen screws 5 and 6 until the front panel can
be lifted out in the forwards direction.

Take the Peltier base plate. Connect the power supply in
the bottom of the sample chamber to the connector on
the underside of the base plate. Place the base plate in the
sample chamber. Replace screws 1 to 4. Take the Peltier
front panel and slot into place before retightening screws
5 and 6.