Jenway Genova Nano German User Manual

Page 95

Advertising
background image

6. Connect a suitable length of sipper pump tubing to the
external outlet port.

7. Insert one end of the capillary tube into the sipper pump
tubing, as shown.

8. Feed the other end through the inlet port located on the
inside of the bulkhead.

9. Fit the sipper probe and secure using the thumbscrew.

10. Carefully feed the tubing through the sipper probe,
allowing sufficient length for it to pass into a suitable
receptacle.

When the sipper accessory has been fitted and the tubing
has been connected the instrument will look like this.

18.2.3.4 Combined sipper Peltier pump

Refer to section 18.2.3.3 for more details.

18.3

USING THE ACCESSORIES

18.3.1 Automatic 8 cell turret

When the automatic 8 cell turret is in use the 8 cell turret
icon is displayed in the bottom right hand corner of the
screen. The current cell position is displayed adjacent to
the 8 cell turret icon. The 0 position should always be used
for the zero calibration sample.

To perform measurements using the automatic 8 cell turret, insert the cuvettes containing the samples
into turret positions 1 to 7. Insert the cuvette containing the blank solution into turret position 0. Enter
the required measurement mode and set up the required measurement parameters. Press the key below
the calibrate to zero icon. The instrument will automatically move the turret around to position zero to
perform the measurement. Once the calibration is complete the measure sample icon will appear and
the turret will return to its original starting position.