Martel Electronics 3001 User Manual

Page 24

Advertising
background image

d)

Press the numeric key corresponding to the custom RTD curve to be used.

e)

At the "SET(1)/RECALL(2)" prompt, press

to recall the custom RTD curve

coefficients.

f)

To use a different custom RTD curve, press the

key twice to display the

USR_DEF selection prompt.

The USR_DEF function of the 3001 uses the Calendar-Van Dusen equation for sourcing
and measuring custom RTD's. The C coefficient is only used for the subrange -260 to 0
degrees Celsius. Only the A and B coefficients are needed for the subrange 0 to 630
degrees. The R0 value is the resistance of the probe at 0 degrees Celsius.

All 5 of the custom RTD curves are set to PT385 at the factory, as shown in Table 2.

Table 2 - Default Custom RTD Coefficients

Curve

Subrange

R0

Coefficient A

Coefficient B

Coefficient C

1

0 to 630

100

3.9083e10-3

-5.775e10-7

0

2

-260 to 0

100

3.9083e10-3

-5.775e10-7

-4.183e10-12

3

0 to 630

100

3.9083e10-3

-5.775e10-7

0

4

-260 to 0

100

3.9083e10-3

-5.775e10-7

-4.183e10-12

5

0 to 630

100

3.9083e10-3

-5.775e10-7

0

Table 3 shows the coefficients for RTD types PT391 and PT392. The C coefficient is only
used for temperatures below 0 degrees Celsius.

Table 3 - Other Common RTD Coefficients

RTD Type

R0

Coefficient A

Coefficient B

Coefficient C

PT392

100

3.9848e10-3

-5.87e10-7

-4.0e10-12

PT391

100

3.9692e10-3

-5.8495e10-7

-4.2325e10-12

4.6

Standard Platinum Resistance Thermometer (SPRT) Coefficients

The SPRT function of the 3001 uses ITS-90 standard coefficients as a basis for measuring
a SPRT. The five custom coefficients are entered as deviations from the standard
coefficients, and as such, all of them are set to zero at the factory.

The coefficients A- and B- represent the A

4

and B

4

coefficient, obtained when the SPRT is

calibrated at the triple points of argon, mercury and water. This covers the 83.8058K to
273.16K subrange. Coefficients A, B and C can represent different coefficients based on
which subranges of the SPRT has been calibrated. For example, if the 273.15K to
933.473K subrange was used, A, B and C would represent A

7

, B

7

and C

7

whereas if the

273.15K to 692.67K subrange was used, A and B would represent A

8

and B

8

and C=0.

To enter the deviation coefficients for a custom SPRT:

a)

Select RTD measure mode as described in the preceding section.