Martel Electronics 3001 User Manual

Page 66

Advertising
background image

OPER?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the present operate/standby mode setting.

Parameter:

<None>

Response:

<value>
where <value> is "1" for operate mode and "0" for standby mode

Example:

OPER?

1

This example indicates that the 3001 is in operate mode.

*OPT?

IEEE-488

RS-232

Sequential

Overlapped

This command returns a list of the installed hardware and software options. This
command is reserved for future use.

Parameter:

<None>

Response:

<text string>
where <text string> is "0" if no options are installed, or a list of installed
options separated by commas

Example:

*OPT?

0

The example indicates that no options are installed.

OUT

IEEE-488

RS-232

Sequential

Overlapped

This command sets the output mode and value of the 3001. To source a temperature,
select the desired mode and sensor parameters first with the TSENS_TYPE, RTD_TYPE,
and TC_TYPE commands. Use the multiplier prefixes k for kilo, m for milli, and u for
micro with the OUT command units, as desired. The unit may be omitted if the output
mode is not to be changed.

Parameter:

<value> <units>
where <value> is the output value to be generated

and where <units> is one of the following, with a possible prefix
multiplier (k, m, or u) as described above:

V

DC volts

A

DC current

OHM

Resistance

CEL

Temperature in Celsius

FAR

Temperature in fahrenheit

Response:

<None>

X

X

X

X

X

X

X

X

X