7.11

Precision Temperature Measurement with IBP-2 Probe

Figure 30 - Precision Temperature Measurement with IBP-2 Probe

1.

With the IBP-2 probe and the corresponding custom coefficients, the total system
error is 0.03°C.

2.

Disconnect any test leads from external devices.

3.

Select RTD input on the primary display as described in section 4.3. Select the user
defined curve containing the custom coefficients for the IBP-2. If custom coefficients
have not yet been entered for the IBP-2, follow the instructions in section 4.5 to do so.

4.

Connect the probe as shown in figure 30.

36