2.2 Primary Input/Output Terminals

Figure 2 describes the primary input/output terminals in detail. The display and controls
for these terminals are described in the next section.

Figure 2 - Primary Input/Output Terminals

Item

Name

Description

1

VOLTS

DC voltage output terminals. See notes 1 and 2 below.

2

mA

DC current output terminals. See notes 1 and 2 below.

3

RTD/O OUTPUT

Two wire RTD and Ohms output terminals. See notes 1
and 2 below.

4

TC INPUT/OUTPUT

Thermocouple input and output terminals. These termi-
nals accept a miniature polarized thermocouple plug with
flat in-line blades spaced 7.9 mm (0.312 in.) center to cen-
ter.

5

4W RTD/O INPUT

Four wire RTD and Ohms input terminals. See notes 1
and 3 below.

6

Pressure module input connector.

Note 1:

These terminal binding posts are made of a special copper alloy to reduce thermal
EMF's. They support the use of either discreet wires or standard banana plugs, and the
HI/LO pairs are spaced for standard dual banana plugs.

Note 2:

Caution. Do not exceed a maximum of 100 volts to chassis ground.

Note 3:

Caution. Do not exceed a maximum of 20 volts to chassis ground.