Martel Electronics 3001 User Manual

Page 78

Advertising
background image

TSENS_TYPE?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the present temperature mode, thermocouple (TC) or Resistance
Temperature Detector (RTD).

Parameter:

<None>

Response:

<value>
where <value> is one of the following:

TC

Thermocouple

RTD

Resistance Temperature Detector

Example:

TSENS_TYPE?

TC

This example indicates that the present temperature mode is
thermocouple.

*TST?

IEEE-488

RS-232

Sequential

Overlapped

This command runs a series of self-tests and returns a "0" for pass or a "1" for fail. If any
faults are detected, they are logged into the fault queue where they can be read by the
FAULT? query.

Parameter:

<None>

Response:

<value>
where <value> is one of the following:

0

failed self test

1

passed self test

Example:

*TST?

1

This example runs the self test and indicates that it passed.

VAL?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the last values for the isolated and primary measurements, in that
order.

Parameter:

<None>

Response:

<iso-value>,<iso-units>,<primary-value>,<primary-units>
where <iso-value> is the present isolated measurement expressed in
scientific notation.

and where <iso-units> is one of the following:

V

DC volts

A

DC current

OVER

measurement is over or under range

or

one of the pressure units listed with the
ISO_PRES_UNIT? command

and where <primary-value> is the present primary measurement

X

X

X

X

X

X

X

X

X

74