12.8 Thermocouple Specification, Output and Input

Absolute Uncertainty,

tcal ±5 °C, ±(°C)

1,2

Range (°C)

Output/Input

TC Type

Minimum

Maximum

90 days

1 year

B

600 °C

800 °C

0.42 °C

0.46 °C

800 °C

1550 °C

0.39 °C

0.39 °C

1550 °C

1820 °C

0.44 °C

0.45 °C

C

0 °C

150 °C

0.25 °C

0.30 °C

150 °C

650 °C

0.21 °C

0.26 °C

650 °C

1000 °C

0.23 °C

0.31 °C

1000 °C

1800 °C

0.38 °C

0.50 °C

1800 °C

2316 °C

0.63 °C

0.84 °C

E

-270 °C

-100 °C

0.38 °C

0.50 °C

-100 °C

-25 °C

0.16 °C

0.18 °C

-25 °C

650 °C

0.14 °C

0.16 °C

650 °C

1000 °C

0.16 °C

0.21 °C

J

-210 °C

-100 °C

0.20 °C

0.27 °C

-100 °C

-30 °C

0.18 °C

0.20 °C

-30 °C

760 °C

0.14 °C

0.17 °C

760 °C

1200 °C

0.18 °C

0.23 °C

K

-270 °C

-100 °C

0.25 °C

0.33 °C

-100 °C

-25 °C

0.19 °C

0.22 °C

-25 °C

120 °C

0.14 °C

0.16 °C

120 °C

1000 °C

0.19 °C

0.26 °C

1000 °C

1372 °C

0.30 °C

0.40 °C

L

-200 °C

-100 °C

0.37 °C

0.37 °C

-100 °C

800 °C

0.26 °C

0.26 °C

800 °C

900 °C

0.17 °C

0.17 °C

N

-270 °C

-100 °C

0.33 °C

0.40 °C

-100 °C

-25 °C

0.20 °C

0.24 °C

-25 °C

410 °C

0.16 °C

0.19 °C

410 °C

1300 °C

0.21 °C

0.27 °C

R

-50 °C

250 °C

0.58 °C

0.58 °C

250 °C

400 °C

0.34 °C

0.35 °C

400 °C

1000 °C

0.31 °C

0.33 °C

1000 °C

1767 °C

0.30 °C

0.40 °C

S

-50 °C

250 °C

0.56 °C

0.56 °C

250 °C

1000 °C

0.36 °C

0.36 °C

1000 °C

1400 °C

0.30 °C

0.37 °C

1400 °C

1767 °C

0.35 °C

0.46 °C

T

-270 °C

-150 °C

0.51 °C

0.63 °C

-150 °C

0 °C

0.18 °C

0.24 °C

0 °C

400 °C

0.13 °C

0.16 °C

U

-200 °C

0 °C

0.56 °C

0.56 °C

0 °C

600 °C

0.27 °C

0.27 °C

1. Does not include thermocouple wire error.