Measurement Commands

Command

Description

ISO_MEAS

Sets the isolated input measurement type.

PRES?

Queries the attached pressure module for its manufacturer and serial num-
ber.

PRES_MEAS

Changes the operating mode of the primary display to pressure measure-
ment.

RTD_MEAS

Changes the operating mode to RTD measurement.

TC_MEAS

Changes the operating mode to thermocouple measurement.

VAL?

Returns the last values for the isolated and primary measurements, in that
order.

ZERO_MEAS

Zeros the pressure module, TC mV, or RTD ohms.

ZERO_MEAS?

Returns the zero offset for the pressure module, TC mV, or RTD ohms.

RS-232 Operating Mode Commands

Command

Description

LOCAL

Puts the 3001 into the local state and disables lockout.

LOCKOUT

Puts the 3001 into the lockout state. This command duplicates the IEEE-488
LLO (Local Lockout) message.

REMOTE

Puts the 3001 into the remote state. This command duplicates the IEEE-488
REN (Remote Enable) message.

Status Commands

Command

Description

FAULT?

Returns the most recent error code in the 3001 error queue, and then
removes that error code from the queue.

10.3 Error Code Listing

Error Number Message Class

Description

1

DDE

Error queue overflow.

100

EXE

OPER or STBY was received when the 3001 is in measure
mode or thermocouple source mode.

101

CME

A non-numeric entry was received in a field that should con-
tain a numeric entry.

102

EXE

The numeric field exceeds 10 characters.

103

CME

Invalid units name or prefix.

104

EXE

An attempt to enter RTD source mode was made when SPRT
is selected, or to select SPRT when RTD source mode is
selected.

105

EXE

Entry is above upper limit for the selected output range.

106

EXE

Entry is below lower limit for the selected output range.