RANGELCK?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the DC voltage range lock status.

Parameter:

<None>

Response:

<value>
where <value> is one of the following:

ON

DC voltage range lock is on

OFF

DC voltage range lock is off

Example:

RANGELCK?

OFF

This example indicates that the range lock is off.

REMOTE

IEEE-488

RS-232

Sequential

Overlapped

This command places the 3001 into the remote state. It duplicates the IEEE-488 REN
(Remote Enable) message. When the 3001 is in the remote state, but not locked out,
only the LOCAL key is active. If the front panel is also locked out, no front panel keys are
active; see the LOCKOUT command. To unlock the front panel, use the LOCAL
command, or cycle the 3001 power switch.

Parameter:

<None>

Response:

<None>

Example: REMOTE

This example places the 3001 into the remote state.

*RST

IEEE-488

RS-232

Sequential

Overlapped

This command resets the 3001 to the power-up state and holds off execution of
subsequent commands until the reset operation is complete.

A reset action invokes the following commands and values for the primary display:

Command

Value

OUT

0 V

PRES_UNIT

Last selected

RANGE

0.1 V

RTD_TYPE

Last selected

STBY

(No output)

TC_REF

INT

TC_TYPE

Last selected

TSENS_TYPE

Last selected

The isolated display and selections remain as they were last selected.

Parameter:

<None>

Response:

<None>

Example:

*RST

This example resets the 3001, invoking the commands and values
shown above.

X

X

X

X

X

X

X

X