Martel Electronics 3001 User Manual

Page 62

Advertising
background image

ISO_MEAS

IEEE-488

RS-232

Sequential

Overlapped

This command sets the isolated measurement type.

Parameter:

<value>
where <value> is one of the following:

DC10V

measure DC voltage, 10V range

DC100V

measure DC voltage, 100V range

DCI

measure DC current

PRESSURE measure

pressure

Response:

<None>

Example:

ISO_MEAS DCI

This example sets the isolated measurement to DC current.

ISO_PRES_UNIT

IEEE-488

RS-232

Sequential

Overlapped

This command sets the isolated pressure unit.

Parameter:

<value>
where <value> is one of the following:

PSI

pounds per square inch

INH2O4C

inches of water at 4 °C

INH2O20C

inches of water at 20 °C

INH2O60F

inches of water at 60 °F

CMH2O4C

centimeters of water at 4 °C

CMH2O20C

centimeters of water at 20 °C

MMH2O4C

millimeters of water at 4 °C

MMH2O20C

millimeters of water at 20 °C

BAR

bars

MBAR

millibars

KPA

kilopascals

MPA

megapascals

INHG

inches of mercury at 0 °C

MMHG

millimeters of mercury at 0 °C

KG/CM2

kilograms per square centimeter

Response:

<None>

Example:

ISO_PRES_UNIT BAR

This example sets the isolated pressure unit to bars.

X

X

X

X

X

X

58