Examples:

Explanation

OUT 15.2 V

Change to DC volts, output 15.2 V

OUT 1.2 mA

Change to DC current, output 1.2 mA, note units prefix
multiplier m

OUT 5 Ohm

Change to Resistance, output 5 O

OUT 100 CEL

Change to temperature in °C, output 100 °C

OUT 3

No change to output mode, output a value of 3 in the
present units

OUT?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the present output value and units of the 3001.

Parameter:

<None>

Response:

<value>,<units>
where <value> is the present output value

and where <units> is one of the following:

V

DC volts

A

DC current

OHM

Resistance

CEL

Temperature in Celsius

FAR

Temperature in fahrenheit

Examples:

Explanation

OUT?
1.88300E-02,A Present output is 18.83 mA

OUT?
1.23000E+00,V Present output is 1.23 V

OUT?
4.00000E+03,OHM

Present output is 4.0 KO

OUT?
1.0430E+02,CEL

Present output is 104.3 °C