Thermocouple Specification, Output and Input (continued)

Absolute Uncertainty,

tcal ±5 °C, ±(°C)

1,2

Range (°C)

Output/Input

TC Type

Minimum

Maximum

90 days

1 year

XK

-200 °C

-100 °C

0.22 °C

0.22 °C

-100 °C

300 °C

0.12 °C

0.13 °C

300 °C

800 °C

0.19 °C

0.20 °C

BP

0 °C

200 °C

0.42 °C

0.42 °C

200 °C

600 °C

0.32 °C

0.32 °C

600 °C

800 °C

0.39 °C

0.40 °C

800 °C

1600 °C

0.45 °C

0.46 °C

1600 °C

2000 °C

0.57 °C

0.58 °C

2000 °C

2500 °C

0.67 °C

0.80 °C

1. Does not include thermocouple wire error.
2. Absolute Uncertainty is inclusive of any cold juction compensation error, which is not stated

separately.