Thermocouple Specification, Output and Input (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)
1,2
Range (°C)
Output/Input
TC Type
Minimum
Maximum
90 days
1 year
XK
-200 °C
-100 °C
0.22 °C
0.22 °C
-100 °C
300 °C
0.12 °C
0.13 °C
300 °C
800 °C
0.19 °C
0.20 °C
BP
0 °C
200 °C
0.42 °C
0.42 °C
200 °C
600 °C
0.32 °C
0.32 °C
600 °C
800 °C
0.39 °C
0.40 °C
800 °C
1600 °C
0.45 °C
0.46 °C
1600 °C
2000 °C
0.57 °C
0.58 °C
2000 °C
2500 °C
0.67 °C
0.80 °C
1. Does not include thermocouple wire error.
2. Absolute Uncertainty is inclusive of any cold juction compensation error, which is not stated
separately.
83