RTD_MEAS

IEEE-488

RS-232

Sequential

Overlapped

This command places the primary display in RTD measure mode.

Parameter:

<value>
where <value> is one of the following:

CEL

display in degrees celsius

FAR

display in degrees fahrenheit

<None>

display in the last selected temperature unit

Response:

<None>

Example:

RTD_MEAS CEL

This example sets the 3001 to RTD measure mode, displaying in
degrees celsius.

RTD_TYPE

IEEE-488

RS-232

Sequential

Overlapped

This command sets the Resistance Temperature Detector (RTD) sensor type for RTD
source and measure. Normally, before using the RTD_TYPE command to select the RTD
type, use the TSENS_TYPE command to select RTD mode, and afterwards set the output
temperature using the OUT command, if applicable. A change in temperature sensors
sets the output to 0 °C. Note that the SPRT can only be used to measure signals, not
source them.

Parameters:

<value>
where <value> is one of the following:

PT385_100

100-ohm RTD, curve a=0.00385 ohms/ohm/°C

PT385_200

200-ohm RTD, curve a=0.00385 ohms/ohm/°C

PT385_500

500-ohm RTD, curve a=0.00385 ohms/ohm/°C

PT385_1000

1000-ohm RTD, curve a=0.00385 ohms/ohm/°C

PT392_100

100-ohm RTD, curve a=0.003926 ohms/ohm/°C

PTJIS_100

100-ohm RTD, curve a=0.003916 ohms/ohm/°C

CU10

10-ohm RTD, empirical curve

NI120

120-ohm RTD, empirical curve

YSI_400

YSI thermistor curve

OHMS_HIGH

4000 ohms range

OHMS_LOW

400 ohms range

SPRT

standard PRT with user defined error coefficients, only
available for measurement

USR_DEF<x> RTD with user defined custom coefficients, where x is

the curve number and ranges from 1 to 5, i.e.
USR_DEF2 for curve 2

Response:

<None>

Example:

RTD_TYPE PTJIS_100

This example sets the RTD sensor to a 100-ohm type, using the
PT3916 curve (a=0.003916 ohms/ohm/°C). The resistance of 100
ohms refers to the ice point characteristic, the resistance of the RTD at
0 °C (32 °F).