Martel Electronics 3001 User Manual

Page 75

Advertising
background image

STBY

IEEE-488

RS-232

Sequential

Overlapped

This command places the 3001 in standby mode, deactivating the output at front panel
terminals. This command acts the same as pressing the front panel key when in
operate mode.

Parameter:

<None>

Response:

<None>

Example:

STBY

This example disconnects the selected output from the 3001 front
panel terminals. It also indicates Stby on the display.

TC_MEAS

IEEE-488

RS-232

Sequential

Overlapped

This command places the primary display in thermocouple measure mode.

Parameter:

<value>
where <value> is one of the following:

CEL

display in degrees celsius

FAR

display in degrees fahrenheit

<None>

display in the last selected temperature unit

Response:

<None>

Example:

TC_MEAS FAR

This example sets the 3001 to thermocouple measure mode,
displaying in degrees fahrenheit.

TC_REF

IEEE-488

RS-232

Sequential

Overlapped

This command selects the internal temperature sensor or an external reference value for
cold junction compensation of thermocouple source and measurement.

Parameter:

<value>
where <value> is one of the following:

INT

use internal temperature sensor

EXT

use external reference value

Response:

<None>

Example:

TC_REF EXT

This example sets the thermocouple reference to external.