Martel Electronics 3001 User Manual

Page 63

Advertising
background image

ISO_PRES_UNIT?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the isolated pressure unit.

Parameter:

<None>

Response:

<value>
where <value> is one of the following:

PSI

pounds per square inch

INH2O4C

inches of water at 4 °C

INH2O20C

inches of water at 20 °C

INH2O60F

inches of water at 60 °F

CMH2O4C

centimeters of water at 4 °C

CMH2O20C

centimeters of water at 20 °C

MMH2O4C

millimeters of water at 4 °C

MMH2O20C

millimeters of water at 20 °C

BAR

bars

MBAR

millibars

KPA

kilopascals

MPA

megapascals

INHG

inches of mercury at 0 °C

MMHG

millimeters of mercury at 0 °C

KG/CM2

kilograms per square centimeter

Example:

ISO_PRES_UNIT?

BAR

This example indicates that the isolated pressure unit is bars.

LOCAL

IEEE-488

RS-232

Sequential

Overlapped

This command puts the 3001 into the local state, clearing the remote state (see the
REMOTE command) and the front panel lockout state (see the LOCKOUT command). It
duplicates setting the IEEE-488 REN line to false.

Parameter:

<None>

Response:

<None>

Example:

LOCAL

This example puts the instrument into the local state, clearing the
remote state and front panel lockout state, if enabled.

X

X

X

X

X