Martel Electronics 3001 User Manual

Page 64

Advertising
background image

LOCKOUT

IEEE-488

RS-232

Sequential

Overlapped

This command puts the 3001 into the lockout state when in remote control (see the
REMOTE command). In this state, no local operation is allowed at the front panel,
including the LOCAL key. To clear the lockout condition, use the LOCAL command. This
command duplicates the IEEE-488 LLO (Local Lockout) message.

Parameter:

<None>

Response:

<None>

Example:

LOCKOUT

This example puts the instrument into the lockout state. No front panel
controls can be used.

LOOP_POWER?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the isolated DC current input 24V loop power status.

Parameter:

<None>

Response:

<value>
where <value> is ON or OFF

Example:

LOOP_POWER?

OFF

This example indicates that the isolated DC current input 24V loop
power is turned off.

LOOP_POWER_OFF

IEEE-488

RS-232

Sequential

Overlapped

This command disables the isolated DC current input 24V loop power.

Parameter:

<None>

Response:

<None>

Example:

LOOP_POWER_OFF

This example disables the isolated DC current input 24V loop power.

LOOP_POWER_ON

IEEE-488

RS-232

Sequential

Overlapped

This command enables the isolated DC current input 24V loop power.

Parameter:

<None>

Response:

<None>

Example:

LOOP_POWER_ON

This example enables the isolated DC current input 24V loop power.