Analog output wiring diagrams – Badger Meter M-Series M4000 User Manual

Page 14

Advertising
background image

14

Terminal 1 Input 1 (+) - Input -
Functions: reset, positive zero return.

Terminal 2 Output 1 (+) Programmable passive output to
Badger

®

external counter.

Functions: forward pulse, frequency output, preset output,
fl ow set point, error alarm, fl ow direction
Active Output to external counter.

Terminal 3 Output 1 (+) ProgrammableTransistor Output
(open collector) - Passive Output to Badger external counter.
Functions: forward pulse, AMR pulse, fl ow set point, error
alarm, empty pipe, fl ow direction.
Active Output to external counter.

Terminal 4 Common fi eld ground.

Terminal 5 Analog Output.

Terminal 6 Ground from external counter device connected
to terminal 5.

Auxiliary Input Wiring Diagram

Function: Reset
Positive

Zero

Return

Input 1 (+)

1

Common 4

5-24VDC

_____________________________________________

Analog Output Wiring Diagrams

• Analog Output (Loop voltage = 18VDC Sourced, 750 MAX)

Function: 4-20mA
0-20mA
0-10mA

mA Analog Output (+) 5

Common 6

_____________________________________________

Functions:

Forward (only) Pulse

AMR

Pulse

Flow

Set

Point

Error

Alarm

Empty

Pipe

Flow

Direction

Output 1 (+)

3

Common 4

02

01

+

External Counter
Badger Meter ER8
or AMR Device

}

Programmable

• Output 1

Transistor Output (open collector)

24VDC

Max

.5W

Max


Passive Output Wiring

_____________________________________________

• Output 2

Transistor Output (open collector)

24VDC

Max

.5W

Max


Passive Output

Functions:

Reverse (only) Pulse

Frequency

Output

Preset

Output

Flow

Set

Point

Error

Alarm

Flow

Direction

}

Programmable

Output 2 (+)

5

Common 6

02

01

+

External Counter
Badger Meter ER8
or AMR Device

External Disconnect

Position this device in an accessible location.

Position and identify the disconnect device so as to provide
safe and easy operation.

Label the disconnect device as being for the Mag Meter.

Install an external disconnect switch or circuit breaker that
meets local standards.

AC Power Wiring
For AC power use three wire, sheathed, cable with cable
diameter of 18 AWG (not supplied).

AC wire insulation temperature class must not exceed the
maximum ambient temperature of its location (typical, 185°F,
85°C).

Use conduit and conduit fi ttings (not supplied) that are rated
for Class I, Div 1 hazardous locations. To maintain a NEMA
4X rating, use watertight fi ttings that are rated NEMA 4X or
better.

To prevent accidents connect main power only after all
other wiring has been completed.