Badger Meter M-Series M4000 User Manual

Page 15

Advertising
background image

15

The amplifi er is a microprocessor device. It is important
that the power supply be as “clean” as possible. Avoid using
power lines that feed heavy loads; pumps, motors, etc.
If dedicated lines are not available, a fi ltering or isolation
system may be required.

AC wiring is the same for meter mount and remote mount
amplifi ers.

1. Remove AC Wires Port screw from amplifi er connections

chamber.

2. Lay out AC cable and conduit to amplifi er.
3. Place cable in conduit.
4. Strip AC cable back 2 inches (50mm).
5. Strip AC wires back ¼ inch (6mm).
6. Attach AC wires to amplifi er terminal.













Amplifier

AC wires

and

terminal

Black to L
White to N
Green to G

L

N

G

85-265 VAC

45-65 Hz

7. Connect conduit to amplifi er.
8. Attach chamber cover.

Adjustable Display/Control Card
Because meter positioning sometimes places the amplifi er
display/programming chamber in an awkward position, the
display/control card is adjustable in 90-degree increments.

Disconnect main power to the unit before attempting any
device maintenance.

To re-position or rotate the display/control card in the amplifi er:

1. Remove display chamber cover. Turn the cover

counterclockwise to remove it from the amplifi er. If
necessary, use a strap wrench.

2.

Remove the 2 card screws and washers. (NOTE:
Use a split screwdriver to prevent dropping screw into
enclosure.)

3.

Tilt card up/out approximately 45 degrees at the holding
clips.

4.

Gently pull card down and out from between holding
clips.

5.

Rotate card to appropriate position.

6.

Angle card and position card holes between the holding
clips.

7.

Push card in, between holding clips. Lower card back
into position and attach card with screws and washers.

Advertising