Menu structure, Flow chart _ menu structure – Badger Meter M-Series M4000 User Manual

Page 33

Advertising
background image

33

In the Flow Chart:

1.

Each separate screen is inside a box.

2.

If a screen “branches”, an arrow points right and each screen of the branch is in a box.

3.

When there is text, but it’s not in a box, that signifi es a parameter setting and a short explanation is given of the parameter.

Mark on the chart what parameters need to be set. Note parameter settings on the chart. Program your amplifi er accordingly.

Keep the chart as a reference for other shifts and personnel and to monitor meter performance.

FLOW CHART _ MENU STRUCTURE

Menu Structure

Meter Setup

Pipe Dia.

Enter the flow tube diameter

Detect Factor

Enter the meter calibration factor

Excitation Freq

Select the coil excitation frequency

Calibration

Electric Zero

All calibration factors

Elec Full Scale

are determined at the

Hydr. Zero-Man

factory.

Hydr. Zero-Auto

Empty Pipe

Cal. empty pipe

Empty pipe detection setup.

Cal. full pipe

Enable/Disable

Chg. Password

Password new

Enter 0000 to disable password.

Measurements

Rate Unit

Select flow rate unit

Totalizer Unit

Select total volume unit

FullScaleFlow

Enter full scale flow for the analog output.

LowFlowCutOff

Enter the low flow cut off. Typically 0.2%.

Flow Dir.

Select uni-directional or bi-directional flow.

Damping Factor

Select as needed.