Press +, - or E once.

7. Main Screen appears. Press E again.

8. PASSWORD??? 01 screen reactivates. Input the correct

password.

Set Flow Rate - Unit of Measure and Totalizer Unit of

Measure

1. With Main Screen up, press E.

2. With MAIN MENU 00 active, press + twice

(Measurements) and then E.

3. On the MEASUREMENTS 20 screen press + once to

place cursor at Flow Unit. Press E.

4. From FLOW UNITS 21 press + or - to select a fl ow unit.

Press E.

5. This type of screen shows for two seconds.

6. The screen returns to MEASUREMENTS 20. With the

cursor at Exit this Menu, press + twice to move cursor to
Totalizer Unit. Press E.

7. From TotalizerUnit 22 press + or - to select a totalizer unit.

Press E.

8. This type of screen shows for two seconds.

9. Screen goes back to MEASUREMENTS 20.

Set Full Scale Flow Rate Value
1. Press + three times to move the cursor to FullScaleFlow.

Press E.

2. At FullScaleFlow 23 input the proper fl ow. Select E to

move cursor as described earlier.

3. Screen returns to MEASUREMENTS 20.

SET LOW FLOW CUTOFF
1. Press + four times, to place cursor at LowFlowCutOff.

Press E.