You only need to do this if the function of output one (1) is
to be Fwd Pulse or AMR (50ms pulse.) Refer to page 35,
"Digital Output #1."

5. In PULSES/UNIT 3D input a proper number as

described earlier. Press E.

6. The screen returns to DIG OUTPUT 1 33. Press or

activate + twice, then press E.

This is only for Fwd Pulse. (Refer to page 35, "Digital Output
#1.")

7. Into screen PULSE WIDTH 3E enter a number. Press E.

8. The screen returns to DIG OUTPUT 1 33. Press + three

times, then press E.

SET POINT

This is only for Flow Set Point. (Refer to page 35, "Digital
Output #1.")

1. Into screen SET PT. MIN 3F enter a number as

described earlier. Press E.

NOTE: When the fl ow rate falls below the minimum setpoint

valvue (entered as a percentage of full scale), the output

will activate.

2. The Screen returns to DIG OUTPUT 1 33. Press + four

times, then press E.

3. In SET PT. MAX 3G enter a number as described earlier.

NOTE: When the fl ow rate exceeds the set point value
(entered as a percentage of full scale), the output will
activate.

This is only for Flow Set Point. (Refer to page 35 "Digital
Output #1.)

OUTPUT TYPE
1. Press or activate the letter E to return to screen DIG

OUTPUT 1 33.

2. In DIG OUTPUT 1 33 press + fi ve times, to Output Type.

3. In OUTPUT TYPE 3H press Normally Open or Normally

Closed.

• Select N.O. or N.C. for any output. Refer to page 35,

"Digital Output #1.

• Normally Opened or Closed **Selected** shows

for two seconds.