Do not lift a detector with a forklift via the detector body between
the fl anges. The housing could be dented and/or damage caused
to internal coil assemblies.

NEVER place forklift forks or rigging chains, straps, slings,

hooks or other objects inside or through the detector fl ow

tube for lifting or handling purposes. The isolating liner could

be damaged, rendering the unit inoperable.

Instructions Specifi c to Hazardous Area Installations
These instructions apply to equipment covered by FM
Certifi cate Number 3015930.

1.

The temperature range for fl uids passing through the
detector is -4°F to 248°F (-20°C to 120°C)

2.

The ambient temperature range surrounding the
amplifi er is -4°F to 122°F (-20°C to 50°C)

3.

The ambient temperature range surrounding the junction
box must not exceed 122°F (50°C)

4.

During any installation or repair, perform all procedures in
accordance with the applicable code of practice.

5.

Suitably trained personnel shall perform all installation
or repair procedures.

6.

If the equipment is likely to come into contact with
aggressive substances, it is the responsibility of the
user to take suitable precautions that prevent it from
being adversely affected, thus ensuring that the type of
protection is not compromised.

Aggressive Substances – e.g. acidic liquids or gases
that may attack metals or solvents that may affect
polymeric materials.

Suitable Precautions – e.g. regular checks as part of
routine inspections or establishing, from the material’s
data sheet, that it is resistant to specifi c chemicals.

Additional Information
Certifi cation markings are noted on the product label.
Markings include:

For additional information regarding importation, equipment
installation, equipment repair, equipment return or renewal
parts, please contact:

Badger Meter, Incorporated
P.O. Box 245036
Milwaukee, WI 53222
Telephone: (414) 355-0400
Fax: (414) 355-7499
On the WEB:

www.badgermeter.com

or contact your local

Badger Meter representative

METER DESCRIPTION
Badger

®

Model M-4000 electromagnetic meters are FM

approved for Class I, Div 1 hazardous locations. To achieve
hazardous location ratings electrodes in the fl ow tube are
Intrinsically Safe, designed according to Factory Mutual (FM)
standards. Those standards limit the amount of energy that
can be sent to electrodes to prevent a spark from occurring.

Empty Pipe Detection
Badger Model M-4000 Mag meters are equipped with an
Empty Pipe Detection feature. Empty Pipe Detection is
accomplished by positioning a third electrode close to the
12 o’clock position. Any time this electrode is not covered by
fl uid, for a minimum of fi ve seconds, the meter displays an
Empty Pipe Detection condition, sends out an error message
if desired, and stops measuring to maintain accuracy. When