Det-Tronics U9500 Infiniti Gas Transmitter User Manual

Page 22

Advertising
background image

95-8444

20

10.1

SETUP PROCEDURE

1. Determine the required alarm setpoint levels,

calibration gas concentration, and other settings

such as range (Table 5), auto/man calibration,

normally energized/de-energized relays, latching/

non-latching relays.

CaUTiOn

Changing the range selection will cause the
alarm setpoints and the calibration gas settings
to default to the setting shown in Table 5. Catalytic
combustible, PointWatch infrared, and oxygen
sensors have preset ranges that are automatically
selected by the Infiniti and cannot be changed.

2. Remove the enclosure cover to access the transmitter

display and controls.

CaUTiOn

While in the Setup Mode, the current loop output
will drop to less than 1 mA. For models without
relays, this is the only remote indication that the
unit is not in the Normal operating mode. For
models with relays, the fault relay indicates a
fault. In either case, the transmitter will remain
in the Setup mode until Setup is complete. The
transmitter must be manually stepped through the
sequence below in order to return to the Normal
operating mode.

3. Press and hold the SETUP/ACCEPT button for 1

second using a small screwdriver, then release. This

initiates the Infiniti Setup Mode. Refer to Table 10

to perform setup. Once a programming option has

been accepted, the transmitter will automatically

cycle to the next option.

CALIBRATION

CALIBRATION RECOMMENDATIONS
The Infiniti can accommodate automatic or manual

calibration and for many sensors either of these methods

are effective. Certain sensors, however, require either

automatic or manual calibration only. The sensors and

method of calibration required and/or recommended are

listed in Table 11. Refer to this table before proceeding

with calibration.

Perform Calibration
• When a new system is initially put into service.
• When the sensor is replaced.
• Periodically to verify proper performance for

combustible, toxic and oxygen detectors - typically

at 90 day intervals, however, this frequency

is dependent upon the requirements of each

application. Refer to the PointWatch manual for

calibration recommendations specific to that

device. Refer to Table 12 for the recommended

calibration frequency for sensors manufactured by

Det-Tronics.

Table 9—Setpoint Display Cycle

Display Indicates

Duration and Description

Normal Mode

To enter the Setpoint Display mode, activate the CAL/RESET magnetic switch for 2 to 3 seconds

(hold the Cal magnet to the side of the transmitter enclosure next to CAL/RESET). If “YES

EXT CAL” was selected during the “Setup Procedure,” the external reset input can also be used to

initiate the Setpoint Display Cycle when activated for 2 to 3 seconds.

Low Alarm Setpoint

1.5 seconds

High Alarm Setpoint

1.5 seconds

Auxiliary Alarm Setpoint

1.5 seconds

Calibration Gas Concentration

1.5 seconds

iMPOrTanT

For catalytic sensors only, whenever the type of calibration gas that is used differs from
the gases to be detected, a Conversion K - Factor must be applied to ensure proper
operation. Refer to “Catalytic Sensor Calibration” section in this manual for details.

Normal Mode

Remains in Normal mode until another mode is activated.

Advertising