Det-Tronics U9500 Infiniti Gas Transmitter User Manual

Page 27

Advertising
background image

95-8444

25

10.1

Table 13—Automatic Calibration Procedure

Description

Display

Operator Action

Normal operation/no gas present Indicates the detected gas

concentration

• If the possibility of background gases exists,

purge the sensor with clean air to ensure accurate

calibration.

Initiate calibration – Catalytic,

PIR9400 & all toxic sensors

Sequences through the Setup

settings

Enters the Calibration mode

-

Display shows detected gas

concentration and alternating

message:

“ZERO” “CAL”

• Hold the Cal magnet to the CAL/RESET magnetic

switch for 7 seconds. See Figure 18 for CAL/RESET

switch location. (The external reset input can also be

used to enter the calibration mode if “YES EXT CAL”

was selected during Setup procedure).

New sensor cal – O

2

New O

2

sensor cal – First remove C7065E cover and

select “Zero”.

Existing sensor cal – O

2

No zero gas action necessary.

Zero calibration complete

Display shows detected gas

concentration and alternating

message:

“APLY”

“GAS”

• Apply the calibration gas to the sensor by placing

the calibration cup over the sensor (or installing

the direct inject fitting and hose on the PointWatch

detector) and opening the valve on the calibration

gas cylinder.

New sensor cal – O

2

Select switch to “ON”. Replace C7065E cover.

Span calibration

Display shows rising gas

concentration and alternating

message:

“GAS”

“ON”

• None.

Span calibration complete

Display shows decreasing gas

concentration when gas is removed

and alternating message:

“CAL” “OK”

then

“RMV”

“GAS”

then, if the calibration is successful:

“XXXX SPAN”

• Remove the calibration gas. When the gas level

falls below the lowest alarm setpoint, and if no faults

are present, the transmitter automatically exits the

Calibrate mode.

• A sensitivity reading that can be used to track

sensor life (except PointWatch) is displayed for

seven seconds with the “SPAN” reading and prior

to the transmitter returning to normal operation. Any

reading over 100 indicates that the sensor is good.

• At the successful completion of the calibration, all

outputs and indicators return to normal operation.

• If faults are present, the unit will exit after the

remainder of 10 minutes.

Calibration fault indication

Display shows alternating messages

indicating the fault, then

“RMV”

“GAS”

• If a fault occurs, remove the gas and correct the fault.

After the fault has been cleared, begin calibration

again. See Table 15 - Fault Messages, Explanations

and Corrective Action.

Advertising