Det-Tronics U9500 Infiniti Gas Transmitter User Manual

Page 28

Advertising
background image

95-8444

26

10.1

MANUAL CALIBRATION PROCEDURE
In the Manual Cal mode, the CAL/RESET switch must be

activated using the Cal magnet to accept a reading as a

calibration point. (The external reset input can also be

used to accept a reading if “YES EXT CAL” was selected

during the Setup procedure.)

The Infiniti must be programmed for manual calibration

prior to initiating the manual calibration procedure (refer

to the Setup section). Refer to Table 14 for the Manual

Calibration Procedure.

iMPOrTanT

The Manual Calibration Procedure must be used
for chlorine (Cl

2

) and sulfur dioxide (SO

2

) sensors.

Use the Auto Calibration Procedure for all other
sensors.

CALIBRATION PROCEDURE FOR INFINITI WITH
C7065E OxYGEN SENSOR
There are two calibration procedures for the U9500/

C7065E:

1. Initial/Replacement Sensor Calibration. This

procedure should be used on new product initial

calibration, or any time an existing sensor has

been replaced with a new or different sensor. This

procedure requires removal of the sensor housing.

Live DC circuits will be exposed to the surrounding

environment during this procedure.

2. Routine Calibration. This procedure should be

used for subsequent calibrations after the initial/

replacement sensor calibration procedure has been

performed. This procedure does not require removal

of the sensor housing, and no live circuits will be

exposed.

Before performing either procedure:

1. Ensure that the device is properly installed and

wired.

2. Ensure that the electrochemical cell is installed

properly.

3. Ensure that the U9500/C7065E is powered-up

and displays normal operating mode without fault

indication. In not, clear all faults before attempting

to enter calibration mode.

4. Ensure that the U9500/C7065E has been

programmed for “auto calibration” mode. Manual

calibration mode is not compatible with the C7065E

sensor.

5. For best results, allow the U9500/C7065E to warm

up for one hour before performing the calibration

procedure.

Initial/Replacement Sensor Calibration Procedure
1. Remove the C7065E aluminum sensor cover. Leave

the sensor cell plugged in.

2. Apply the calibration magnet to the U9500 calibration

target location for nine seconds, or until the U9500

display shows “zero-cal”.

3. Move the toggle switch on the electrochemical

sensor cell to the “zero” position. See Figure 19 for

switch location. This will simulate the “zero” or “no-

oxygen” condition necessary to establish a baseline

output calibration level. When the U9500 display

reads “00”, move the toggle switch back to the “on”

position. Place the aluminum sensor cover over the

cell and reinstall.

4. When the U9500 display shows “aply-gas”, the

transmitter will begin to record the sensor output level

in response to the ambient oxygen concentration.

If this is known to be 20.9%, an accurate span

calculation will result. If the ambient O

2

level is

less than 20.9%, it is recommended to apply 20.9%

oxygen using a Det-Tronics O

2

calibration kit.

5. When span calculations are complete, the U9500

display will indicate “cal-OK”, and then “rmv-gas”.

If using a Det-Tronics O

2

calibration kit, remove the

calibration cup and close the regulator valve. No

actions are necessary if ambient air O

2

is used.

6. The U9500 will automatically exit calibration mode

when calibration is complete. The displayed O

2

level

should be 20.9% (17.4 mA current level) if ambient

conditions are normal.

A1472

ZERO SWITCH

Figure 19—Electrochemical O

2

Sensor Cell

Advertising